Key Features

  • Foam backing materials protect sensor and gel from surgical and cleaning fluids

  • The solid adhesive gel provides consistently high quality recordings for long term monitoring with minimal skin reactions

  • All sensors have an high quality Ag/AgCl layer

Product

Recommended Application

Monitoring

Characteristics

Disposable pregelled, non latex,  no PVC, non sterile

Shelf life

24 months unopened

X-Ray Translucent

-

 

Classification and Standards

Classification (MDD Art. 9)

Class I

Classification (CFR 21 870.2360.)

Class II

 

Dimensions

Electrode shape

Round – 50 mm

Electrode size - max L/W [mm]

-

Total area [mm˛]

1.963

Gel area [mm˛]

299

Adhesive area [mm˛]

1.863

 

Materials

Stud

Stainless steel

Label

PET-foil

Backing material

PE-foam

Adhesive

Medical-grade acrylate

Sensor

Silver/siverchloride (Ag/AgCl)

Sponge

PET-foam

Gel

Solid adhesive gel

Release liner

Siliconized Pet-foil (Transparent)

 

Materials Packaging

Pouches, inner layer

Polyethylene (PE)

Pouches, centre layer

Aluminum (Al)

Pouches, outer layer

Paper

Boxes

Cardboard

 

Standard Packaging

Pieces / card

1

Pieces / pouch

30

Pieces / box

1500

Lot number and expiration date on every pouch and box

 

Dimensions Packaging

Box (L/W/H) [cm]

38,5,2/28,5/21,0

 

Biocompatibility

Test - ISO 10993-1

Backing material (adhesive)

Gel

Cytotoxity

Pass

Pass

Skin Irritation

Pass

Pass

Sensitization

Pass

Pass

 

Electrical Values

Units

Typical Value

AAMI Limits

DC-Offset

[mV]

  1.5

100

DC Offset (5 sec after capacitor discharge)

[mV]

 10

100

Recovery Slope

[mV/sec]

  -0.3

± 1

AC-Impedance with 10 Hertz

[Ω]

 100

2000

Internal Noise Test

  [mV]

 10

150

Bias-Tolerance (over 24 h)

[mV]

 10

100